Link to DACQ workstation is down
Problem
If MEG Operators inadvertently shutdown/restart the MEG DACQ Console/Workstation, rather than logging off, or the DACQ reboots unexpectedly, the IHR will send out the following error SMS text and email ...
Link to DACQ workstation is down
Solution
- If powered off completely, switch on the DACQ PC.
- If DACQ is on, but Link is lost - press
Ctrl-Alt-Deleteat least once, possibly twice, to soft boot.
- If DACQ is on, but Link is lost - press
- Allow the DACQ to reboot whilst watching the bootloader/kernal initialisation process for any errors.
- Allow any disk check to take place.
- If errors are seen, make a note/take 'phone image, and contact MEG Support, or email chbh-help[@]contacts ..., or MEGIN as necessary.
- After any advice, reboot DACQ as necessary.
- Once the Desktop/Login environment has loaded, log in.
- Perform a RAP (Restart Acquisition Programs).
Applications -> Neuromag -> Maintenance -> Restart Acquisition Programs
- Move the gantry to an "in use" position e.g. Upright (60)
- Then start Acquisition and Load the latest Tuning file, see Loading a Tuning File, to make sure the correct Tuning file for acquisition is loaded, and to check for any sensor noise.
-
If the sensors are still noisy, more than one RAP might be required and also maybe a soft boot of the electronics
- Check for any errors in the Message Box at the top of the Acquisition control window, or in any of the Acquisition settings e.g. Gantry Position
-
- See Restart Acquisition for full details.
- At some point after sucessful DACQ reboot, the IHR should send out the following cleared SMS text and email ...
Link to DACQ workstation up again