Problem

If MEG Operators inadvertently shutdown/restart the MEG DACQ Console/Workstation, rather than logging off, or the DACQ reboots unexpectedly, the IHR will send out the following error SMS text and email ...

Link to DACQ workstation is down

Solution

  1. If powered off completely, switch on the DACQ PC.
    • If DACQ is on, but Link is lost - press Ctrl-Alt-Delete at least once, possibly twice, to soft boot.
  2. Allow the DACQ to reboot whilst watching the bootloader/kernal initialisation process for any errors.
    • Allow any disk check to take place.
    • If errors are seen, make a note/take 'phone image, and contact MEG Support, or email chbh-help[@]contacts ..., or MEGIN as necessary.
    • After any advice, reboot DACQ as necessary.
  3. Once the Desktop/Login environment has loaded, log in.
  4. Perform a RAP (Restart Acquisition Programs).
    • Applications -> Neuromag -> Maintenance -> Restart Acquisition Programs
  5. Move the gantry to an "in use" position e.g. Upright (60)
  6. Then start Acquisition and Load the latest Tuning file, see Loading a Tuning File, to make sure the correct Tuning file for acquisition is loaded, and to check for any sensor noise.
    • If the sensors are still noisy, more than one RAP might be required and also maybe a soft boot of the electronics

    • Check for any errors in the Message Box at the top of the Acquisition control window, or in any of the Acquisition settings e.g. Gantry Position
  7. See Restart Acquisition for full details.
  8. At some point after sucessful DACQ reboot, the IHR should send out the following cleared SMS text and email ...

Link to DACQ workstation up again