MaxFilter Fine-Calibration Files

If using MaxFilter for any analysis ...

(Right-click link, "Save link as...").

Supine: sss_cal_3140_0_190213.dat

60 Degree: sss_cal_3140_60_190213.dat

  • now the default "sss_cal.dat" file, as from 24th November 2021.

68 Degree: sss_cal_3140_68_190213.dat

If required, the ctc file (which hasn't changed) ...

ct_sparse_triux2.fif (dated 10/02/2018)

From a note re: duty-of-care overnight Acquired data backup ...

If we can't copy the symbolic links,
new ones will need to be made if we have to copy the files back ...

"sss_cal.dat"

Location: /neuro/databases/sss
Points to: sss_cal_3140_60_190213.dat (as from 24/11/2021)
User: neuromag
Group: neuro

symlink "/projects/meg-data/rds/neuro/databases/sss/sss_cal.dat" -> "sss_cal_3140_60_190213.dat"

(was originally pointng to "sss_cal_3140_190213.dat" - a copy of "sss_cal_3140_0_190213.dat")

_______________________________________

"ct_sparse.fif"

Location: /neuro/databases/ctc
Points to: ct_sparse_triux2.fif
User: neuromag
Group: neuro

symlink "/projects/meg-data/rds/neuro/databases/ctc/ct_sparse.fif" -> "ct_sparse_triux2.fif" 

"MaxFilter fine-calibration file does not affect the recorded raw data at all, they are completely independent. Thus, the 60-degrees calibration can be applied to all recordings in seated position, whenever they have been performed."
- 24th November 2021: Jukka Nenonen, Method Development Manager.