MaxShield state unknown
Problem
On 16th July 2018, whilst in the Tools section of the Acquisition program, the following error message was generated.
403 MaxShield state unknown - cannot measure.
Turn MaxShield on or off
Solution
- Restart/Reset the DSP electronics, see Resetting the electronics.
- Perform a RAP (Restart Acquisition Programs).
- Load the latest Tuning file, see Loading a Tuning File.
See Restart Acquisition for full details.
- If the error message is still not cleared, email MEGIN Support (support [at] megin.com)