MEG Intercom faulty

Problem

On a couple of occasions, pressing "11", as usual, on the intercom produced an atypical beep, then a number of rapid beeps were heard and communication to the MSR failed.

Solution

The MIC SPEAKER PHONE needs to be power cycled.

ODA Alarm

  • Open up the rear door of the Electronics cabinet.
  • Unplug the intercom power cable from the unit located at the bottom of the cabinet (as indicated).
  • Replace the cable after a few seconds.
  • Attempt to open the intercom by pressing "11".