EyeLink Settings
Screen Dimensions:
- Width: 830mm - Width of the viewable portion of the projected Stim PC monitor image.
- Height: 470mm - Height of the viewable portion of the projected Stim PC monitor image.
NOTE: These values may change, depending on the position of the projector screen relative to the projector mirror.
Stimulus paradigm will also determine the screen position - use tape on the MSR floor to help with projector screen placement.
Display Resolution of Stim PC monitor:
- Width: 1920 pixels
- Height: 1080 pixels
This setting only needs to be changed, in FINAL.INI, if a different display resolution is used to the one above.
Eye-to-Screen Distance:
- 1000mm - Distance from the eye to the top of the viewable portion of the projected Stim PC monitor image.
- 1050mm - Distance from the eye to the bottom of the viewable portion of the projected Stim PC monitor image).
NOTE: These values may change, depending on the position of the projector screen relative to the gantry chair/participant. e.g. 1500mm.
Some Operators also measure Distance from the eye to the center of the viewable portion of the projected Stim PC monitor image.
EyeLink lens defualt setup:
- 35mmm - Used by most Operators. 50mm lens is available on the MSR shelving if required.
- Swap back to 35mm after use.
- "I put the eye tracker very close to the participant, so smaller lens will give me a wider angle to cover the whole visual fields" Yali Pan.
Camera-to-Screen Distance:
- 150mm - Only applicable in "Remote" mode (which is not used), so can safely be ignored.